Low coefficient of expansion and thermal stability |
|
Water absorption and porosity is low and hence long term dimensional stability is good |
|
It is fine-grained and the density & hardness is suited for dimensional metrology |
|
Flexural strength, Compressive strength, Modules of rupture and Tensile stress |
|
Totally corrosion free, granite does not rust nor corrode or react chemically with other substances |
|
Non Magnetic and non-conductive |
|
Non-reflective matt finish; Hand lapped for maximum accuracy |
|
High flatness accuracy |
|
Ideal for guideways for measuring instruments and air bearing surfaces |
|
High bending strength |
|
Insensitive to mechanical overloading |
|
All calibration equipment is to national and international laboratories standards |
|
PRECISION GRANITE SURFACE PLATES |
SPECIFICATIONS
|
FLATNESS TOLERANCE IN µm FOR ACCURACY GRADE
|
SIZE OF
PLATE (in mm)
|
AS PER DIN 876:84
|
AS PER IS 7327-2003
|
WEIGHT
in Kgs.
(Approx.)
|
|
GRADE "00"
|
GRADE "0"
|
GRADE "1"
|
GRADE "2"
|
GRADE "3"
|
|
250 x 250
|
3
|
3.5
|
7
|
15
|
30
|
10
|
300 x 300
|
3
|
3.5
|
7
|
15
|
30
|
22
|
400 x 250
|
3
|
4.0
|
8
|
16
|
32
|
20
|
400 x 400
|
3
|
4.5
|
9
|
17
|
34
|
40
|
500 x 300
|
3
|
4.5
|
9
|
17
|
34
|
35
|
630x 400
|
4
|
5.0
|
10
|
20
|
39
|
65
|
630 x 630
|
4
|
5.0
|
10
|
21
|
42
|
100
|
800 x 500
|
4
|
5.0
|
10
|
21
|
42
|
120
|
1000 x 630
|
4
|
6.0
|
12
|
24
|
49
|
320
|
1000 x 1000
|
4
|
7.0
|
14
|
28
|
56
|
500
|
1200 x 800
|
4
|
7.0
|
14
|
28
|
56
|
475
|
1600 x 1000
|
5
|
8.0
|
16
|
33
|
66
|
1100
|
2000 x 1000
|
6
|
9.5
|
19
|
38
|
75
|
1650
|
|
|
Application
Surface plates are used as reference planes
for dimensional inspection and calibration of mechanical artifacts.
This plate acts as a datum line or surface. So, the flatness of the
plate is very important. Dolerite is the perfect material for metrology,
because of the inherent characteristics mentioned above. The definition
of flatness is, ‘all points on the surface being contained within two
parallel planes, the base plane and the roof plane’. This flatness
measurement commonly carries a tolerance and may include a grade
designation. Flatness tolerance of the surface plate is provided in the
standards with different grades based on flatness.
GMT Metrology Granite Surface Plates meet
the DIN and IS Standards specified in the table above. Granite Surface
Plates are manufactured to precision standards using the highest quality
granite.
Accuracy
Rigorous quality control inspection is
carried out at all stages. Final inspection is carried out at a High
Precision Computerised Electronic Level calibrated by an NABL accredited
lab at regular intervals..
Accuracy under Load
Granite Surface Plates are capable of supporting a total normal load of 25 kgs. per 0.1 sq.mtr.
Certificate
A certificate of accuracy is supplied with each plate as per standards specified.
Technical Data
Granite Surface Plates are manufactured in
five grades. The typical application of these five grades of plates is
as under:
GRADE 00 as per DIN : 876 : 84 is of laboratory grade. It is specified for precision measurement in
gauge rooms and metrology laboratories for calibration masters.
GRADE 0 is of inspection grade. It is specified for general work in quality control.
GRADE 1 is of tool room grade. It is specified for production checking work through the shop.
GRADE 2 & 3
are intended for use in shops for general inspection. The choice of
Grade 2 & 3 will depend on the application needs of the customer.
MATERIAL
Granite Surface Plates are made from select
granite with uniform distribution of light and dark constituents. In
such granite the minerals are evenly distributed resulting in a
homogenous appearance.
The following are the representative properties of our granites : |
Texture |
Fine |
Specific Gravity |
3.08 g/cm3 |
Hardness |
Scleroscope over 90 |
Modules of Elasticity |
9.0 to 12.0 x 105kg/cm2 |
Moisture of absorption |
Nil |
|
FLATNESS TOLERANCE
The flatness deviation of a local area of 250 mm x 250 mm of the working surface shall not exceed : |
1
|
3.0 µm for plates of Grade 00 |
2
|
3.5 µm for plates of Grade 0 |
3
|
7.0 µm for plates of Grade 1 |
4
|
15.0 µm for plates of Grade 2 |
5
|
30.0 µm for plates of Grade 3 |
|
GRANITE SURFACE PLATE STANDS
These are fabricated of heavy angle iron and
reinforced to ensure rigidity. The stands are designed to withstand
the weight of the granite surface plates and are
supplied with 3 leveling screws for leveling the granite surface
plates.
OPTIONAL ACCESSORIES
Plate will be supplied with a suitable
Rexine Cover to protect both the working surface and side faces of the
plate at extra cost. |
|